3.2 Status register (0x00, r)
Bit Name
Function
3
FLASH BUSY FLAG
0 = Flash ready 1 = Flash busy
2
MEMORY FULL
0 = Data Memory is not full
1 = Data Memory has been filled
(can be reset via SWR_RESET or FLAG CLEAR bit of CONTROL
REGISTER)
1
MEMORY EMPTY
0 = Data Memory is not Empty 1 = Data Memory is Empty
0
USB TYPE
0 = Full Speed (USB 1.1) 1 = High Speed (USB 2.0)
3.3 Control register (0x01, r/w)
Bit
Name
Function
Default
Value
15
OUTP LEMO LOGIC
LEVEL
sets the OUTP logic level when OUTP MODE is "PROGRAMMABLE LEVEL"
0
13,14
OUTP LEMO MODE
CONFIG
selects the signal to present on OUTP front panel connector
00 = MEMORY FULL (active high); default
01 = PEAK DETECT (active high when stretcher detects one peak)
10 = PROGRAMMABLE LEVEL (Logic level set by OUTP LEMO LOGIC
LEVEL bit)
11 = PULSE MODE (Programmable width active high pulse when PULSER
register is written)
00
12 SLIDING
SCALE
MODE SELECT
selects sliding scale type:
0 = Random mode (default) 1 = RAMP mode
0
11
DAC TEST MODE
ENABLE
0 = DAC test mode disabled
1 = DAC test mode enabled. Sliding scale disabled and and the sliding scale
DAC is set with the DAC register value (see 0) – only for test purpose
0
10 RESERVED
-
9 ACQUISITION
MODE
0 = External Gate mode 1 = Auto Gate mode (on discriminator threshold)
1
8 PUR
INPUT
ENABLE
0 = PUR input sensing disabled 1:= PUR input sensing enabled
1
7 SOFTWARE
CONVERSION
MODE ENABLE
0 = Software conversion mode disabled
1 = Software conversion mode enabled (the control FPGA auto generates
conversion requests at a fixed periodic rate – only for test purpose)
0
6 CONVERSION
ENABLE
enables ADC data storage into memory buffer for readout
0 = ADC data storage disabled 1 = ADC data storage enabled
0
5
FLAG CLEAR
0= Rearms Memory Full Flag 1= Forces Memory Full Flag Clear (see § 3.2)
0
4 SLIDING
SCALE
ENABLE
0 = Sliding Scale disabled 1 = Sliding Scale enabled
1
3..1
ADC SWR RATE
Selects ADC Conversion Rate when ADC Software Conversion is enabled
0 = 600 S/s
1 = 70 KS/s
2 = 140 KS/s
3 = 250 KS/s
4 = 420 KS/s
5 = 635 KS/s
6 = 850 KS/s
7 = 1 MS/s
7
0 Timers/scaler
enable
FLAG
0 = Scaler/Timers disabled 1 = Scaler/Timers enabled
0