BYK LC 4
Operation
41
4.12
Reset
The following parameters are stored in the sensor module. The reset function
resets them to the delivery condition.
Note
The calibration data is lost when the measuring parameters are reset. Reca-
librate after performing a reset. Stored data are retained.
Parameter
Sensor module, TetraCon V
Adjustable values
Delivery
condition
Measuring mode
, Spec. resistance, Sal, TDS
,
with adjusted cell
constant
Cell constant, to be
calibrated
0.450 ... 0.500
0.475
Cell constant, to be
adjusted
0.450 ... 0.500
0.475
Temperature
compensation
nLF, Lin, Off
nLF
linear TC
0.001 ... 3.000 %/K
2.000 %/K
Tref
20 °C, 25 °C
25 °C