Operation
BYK LC 4
24
4.4.2
Determining the cell constant (calibration in the control stan-
dard)
Note
This method of automatically determining the cell constant by calibration in
the 0.01 mol/l KCL control standard solution can only be used for measuring
cells with cell constants in the range 0.450 ... 0.500 cm
-1
(sensor module
TetraCon V).
This is how you can determine the cell constant:
Note
You can return to the measuring mode with
<
g
>
.
The displayed calibrated cell constant is used immediately.
Note
You can terminate the calibration with
<
E
>
.
The data of the terminated calibration is not used.
1
Open the setting menu for the cell constant with
<
C
>
.
The display shows the setting last selected.
2
If necessary, press
<
"_
>
repeatedly until the
CAL
display appears.
3
If necessary, switch between the cell constant last calibrated, the
date of the calibration and the calibration temperature with
<
ß
>
.
4
To calibrate:
Immerse the sensor module in the control standard solution.
5
Start the measurement with
<
C
>
.
The calibrated cell constant is displayed.
The unit of the measured value display flashes (AutoRead) as long
as the measured value is not stable.
ESC
TC
Sal
4
mg/l
Lin
Tref20
Tref20
Tref25
nLF
TDS
MEM
C
mS/cm
RCL
S
SAL
O
2
RESET
Cell
LAP
mbar
STO
LoBat Day.Month h:min min:s
Ident
Lap
CLR
M cm
k cm
K
A L
C
CAL
Cell
4 6 8
0
1
cm
CAL
TC
ESC
CellTC
Sal
4
mg/l
Lin
Tref20
Tref20
Tref25
nLF
TDS
MEM
C
mS/cm
RCL
S
SAL
O
2
RESET
Cell
LAP
mbar
STO
LoBat
h:min min:s
Ident
Lap
CLR
M cm
k cm
1
cm
K
A L
C
CAL
Day.Month
1 3 0 5
CAL
TC
<
ß_
>
ESC
Day.Month
1
CellTC
Sal
4
mg/l
Lin
Tref20
Tref20
Tref25
nLF
TDS
MEM
mS/cm
RCL
S
SAL
O
2
RESET
Cell
LAP
mbar
STO
LoBat
h:min min:s
Ident
Lap
CLR
M cm
k cm
1
cm
K
A L
C
CAL
2 5 7
CAL
TC
C
<
ß_
>
Summary of Contents for LC 4
Page 6: ...Overview BYK LC 4 6...
Page 12: ...Commissioning BYK LC 4 12...
Page 46: ...Maintenance cleaning disposal BYK LC 4 46...
Page 54: ...Technical data BYK LC 4 54...
Page 60: ...Index BYK LC 4 60...