Operation
BYK LC 4
26
4.5
Setting the temperature compensation
The calculation of the temperature compensation is based on the preset
reference temperature,
Tref20
or
Tref25
(see section 4.6 S
ETTING
THE
REFER
-
ENCE
TEMPERATURE
).
You can select one of the following temperature compensation methods:
!
Nonlinear temperature compensation (
nLF
)
!
Linear temperature compensation (
Lin
)
with selectable coefficients of
0.001 ... 3.000 %
/
K
!
No temperature compensation (- - - -)
Note
Select the following temperature compensations given in the table according
to the respective test sample:
Application tips
Measuring solution
Temperature compensation
(TC)
Display
indicator
Natural water (ground
water, surface water
and drinking water)
nLF
according to ISO 7888, EN
27888
(*)
(*) Corresponds to DEV C8
Ultrapure water
nLF
according to ISO 7888, EN
27888
(*)
Other aqueous solu-
tions
Set linear
temperature coefficient 0.001
... 3.000 %/K
Salinity
(seawater)
Automatic nLF according to
IOT
Sal,