Troubleshooting
When a problem occurs on the screw vacuum pump, it must be stopped immediately. Close the suction valve immediately in order to avoid the
ventilation of the vessel through the screw vacuum pump. The pump will then have to be inertised immediately if it cannot be excluded that there is
an explosive gas mixture present inside the pump.
Troubleshooting procedure
–
A problem has been flagged up by a monitoring device during the operation of the pump
–
Close the suction valve immediately and simultaneously supply non-explosive gas through the flushing gas connection, suction side.
–
Non-explosive gas volume flow: NC 400 B
³
100 liters/ min.
–
Inertise for at least 200 seconds, after that shut off the gas supply and the inerting process is completed
–
Switch off the pump and close the valves
Before restarting the pump, the cause of the problem has to be found and resolved. This is particularly important when
TS, LS or SA
have indicated
the problem.
The vacuum pump must be installed in such a way as to prevent any return of gas or liquid drawn medium to get back into the pump when the
pump stops.
WARNING
Risk of electrical shock, risk of damage to equipment.
Electrical installation work must only be executed by qualified personnel.
CAUTION
During operation the surface of the vacuum pump may reach temperatures of more than 90 °C.
Risk of burns!
Let the vacuum pump cool down prior to contact or wear heat protection gloves.
Problem
Possible Cause
Remedy
The vacuum pump does not reach the usual
pressure
The vacuum system or suction line is not
leak-tight
Check the hose or pipe connections for possible
leaks
Wrong motor rotation
Change two phases of motor
Pumping out the system takes too long
In case a mesh screen is installed on the suction
connection:
The mesh screen on the suction connection is
partially clogged
Clean the mesh screen
If cleaning is required too frequently install a fil-
ter upstream
In case an inlet filter is installed on the suction
connection:
The filter on the suction connection is partially
clogged
Clean or replace the inlet filter
Partial clogging in the suction, discharge or
pressure line
Remove the clogging
Long suction, discharge or pressure line with
too small a diameter
Use larger diameter tubing
Internal parts worn or damaged
Repair the vacuum pump (Busch service)
The vacuum pump does not start
The drive motor is not supplied with the correct
voltage or is overloaded
Supply a drive motor with the correct voltage
The drive motor starter overload protection is
insufficient or trip level is too low
Compare the trip level of the drive motor starter
overload protection with the data on the name-
plate
Correct if necessary
In case of high ambient temperature:
Set the trip level of the drive motor starter
overload protection 5 percent above the no-
minal drive motor current
One of the fuses has blown
Check the fuses
Troubleshooting
Page 18
Summary of Contents for 7180551
Page 24: ......
Page 26: ...Temperature sensor TS Temperature sensor TS Page 26...
Page 27: ...Temperature sensor TS Page 27...
Page 28: ...Temperature sensor TS Page 28...
Page 29: ...Temperature sensor TS Page 29...
Page 34: ...Note...
Page 35: ...Note...