Page 41
9.1
Accuracy
Specified accuracy is valid after a 10 minute warm up period in temperature of 23 ± 5 °C.
Uncertainties include long-term stability, temperature coefficient, linearity, load and line regulation and
traceability of factory to national calibration standards.
Accuracies are assigned in % of the set value.
The Specified accuracy is valid for one-year.
Table 4: Resistance accuracy (terminals R4W)
Range
Accuracy
1
Ω
- 400
Ω
0.003 % + 3
mΩ
400
Ω
- 2000
Ω
0.005 %
2000
Ω
- 10000
Ω
0.015 %
Maximal thermoelectric voltage on terminals R4W in resistance range 1
Ω to 2
kΩ is 1
µ
V and in range
2
kΩ to 10
kΩ is 5
µ
V.
Table 5: Resistance accuracy (terminals R2W)
Range
Accuracy
1
Ω
- 2000
Ω
0.005 % + 10
mΩ
2
kΩ
- 200
kΩ
0.005 %
200
kΩ
- 1200
kΩ
0.01 %
The maximum thermoelectric voltage on terminals R2W in resistance range 1
Ω to 2
kΩ is 5
µ
V and in
range 2
kΩ to 1.2
MΩ is 15
µ
V.
Table 6: Pt temperature sensor simulation accuracy
Range
Pt100 (R4W) Pt200 (R4W) Pt500 (R4W) Pt1000 (R4W) Pt10000 (R2W)
-200 to 200
°
C 0.02
°
C
0.02
°
C
0.02
°
C
0.04
°
C
0.04
°
C
200 to 500
°
C 0.03
°
C
0.04
°
C
0.06
°
C
0.1
°
C
0.06
°
C
500 to 850
°
C 0.04
°
C
0.06
°
C
0.15
°
C
0.2
°
C
0.1
°
C
Table 7: Ni temperature sensor simulation accuracy
Temperature range Ni100 (R4W) Ni1000 (R4W) Ni10000 (R2W)
-60 to 300
°
C
0.02
°
C
0.04
°
C
0.04
°
C