Page 35
7.4.4 R
0
setting / reading
R (?) <DNPD>
This Command sets the resistance value to 0 °C.
The setting of R
0
is valid for all types of simulated temperature sensors.
<DNPD>
Represents the resistance value R
0
in Ω.
The limits are shown in chapter 9 “Specifications”. The 1427
confirms the execution with the string „Ok <cr><lf>”.
In case of a query the 1427 returns the set val
ue in Ω.
Example:
„R100 <cr>” sets the value R
0
to 100
Ω (Pt100, Ni100).
After the query „R?<cr>” the decade returns the string „100<cr><lf>”.
7.4.5 Temperature unit setting (T.unit)
U <CPD> { 0 | 1 }
This Command sets the used temperature unit.
Code Function
0
sets degrees of Celsius (
°
C)
1
sets degrees of Fahrenheit (
°
F)
The 1427 confirms execution with the string „Ok <cr><lf>”.
Example:
„U0<cr>” sets
°
C as temperature unit.