Page 28
Table 2:
Maximum deviations (R4W)
Nominal value [Ω]
Max. deviation [mΩ]
1.00000
3.03
2.00000
3.06
5.00000
3.15
10.00000
3.3
20.0000
3.6
50.000
4.5
100.000
6.0
200.00
9.0
500.00
25
1000.0
50
2000.0
100
5000.0
750
10000.0
1500
Connect the standard ohm-meter to the R2W terminals of the resistance decade.
Use four-wire connection for a range up to 10 k
Ω
. Above this point you can use both two-wire and
four-wire connection.
At “SETUP” menu set the parameter “4W<2W” to 0
Ω.
Check resistance value at checkpoints according to table: “Maximum deviations (R2W)”.