![burmeier Regia Instruction Manual Download Page 89](http://html1.mh-extra.com/html/burmeier/regia/regia_instruction-manual_2832795089.webp)
Inspection report
(normalised to
rated value of
mains voltage)
Result: Bed prot.
class II (type BF)
0.1 mA
mA
In case of measured voltage external
conductor - earth
volt:
Inspection report
III. Functional check
OK
Not
OK
Description of
defect
Functional check of the electrical components
What?
How?
End of travel cut-out of the motors
Automatic cut-out
External power supply / handset
No ‘rattling’ when shaken?
Handset: Operating function, lock-
ing function
Perform the test acc. to
Motors
Abnormal noise develop-
ment (rattling, uneven run-
ning)
Strain Relief of mains cable (if
mains cable available)
Mains cable firmly fas-
tened
Functional check of the mechanical components
Joints and pivots
Smooth operation
Grab handle with strap
Securely fixed when load
tested under approx. 75
kg load (hang from it brief-
ly with two hands)
Castors
Moving and braking
Emergency release of the backrest
Test according to instruc-
tion manual
Safety sides
Securely engaged, secure
position, unlocking
Maintenance
Part B: Operator and Technical Personnel
53
Summary of Contents for Regia
Page 2: ...Part A General Information ...
Page 3: ...Regia Standard care bed Regia care bed DSG optional ...
Page 6: ...II ...
Page 30: ...Part B Operator and Technical Per sonnel ...
Page 31: ...Regia Standard care bed Regia care bed DSG optional ...
Page 36: ...IV ...
Page 105: ...Appendix Part B Operator and Technical Personnel 69 ...
Page 106: ...Part C Care staff and residents ...
Page 107: ...Regia Standard care bed Regia care bed DSG optional ...
Page 151: ...Cleaning and Disinfection Part C Care staff and residents 41 ...