80
Troubleshooting with messages
19.3.15
Message
Warning: too low leakage current
Product-status symbol
Bit status of the PDO pH-state
bit 7 = 1
Possible cause
The leakage current of the measuring cell is under the set limit.
The message can only be displayed if the monitoring of the leakage
current has been configured and activated. Refer to chpt.
What to do?
→
As the monitoring of the low limit is not important, make sure
that the parameter
Error triggers
Activation flags
Low ISFET leak. current
is not selected.
19.3.16
Message
Warning: too high leakage current
Product-status symbol
Bit status of the PDO pH-state
bit 6 = 1
Possible cause
The leakage current of the measuring cell is above the set limit.
The message can only be displayed if the monitoring of the leakage
current has been configured and activated. Refer to chpt.
What to do?
→
Acknowledge the warning message: the product continues to
operate.
→
If necessary, replace the measuring cell. Refer to chpt.
→
Check if the high warning limit is consistent. Refer to chpt.
If the high warning limit is not consistent, send the product back
to Bürkert.
English
Type MS01