pH sensor - Diagnostics
Do the calibration procedure:
→
Keep the Type MZ15 handheld calibration-and-cleaning tool upright during calibration procedure
to make sure that both of the conditions are met:
-
The buffer solution flows correctly through the product.
-
The check valves operate correctly.
Step 1/5:
→
To let the buffer solution flow through the product, press the actuator button of the Type MZ15.
→
Select
Next
.
Step 2/5:
→
Select
Input value of buffer solution
.
→
Enter the pH value of the buffer solution.
→
Validate with
.
→
Select
Next
.
Step 3/5:
→
When the pH measurement is stable, select
Next
.
Step 4/5:
There are 3 possible results:
•
The calibration succeeds.
•
The message
Error: Value out of range
is displayed.
•
The message
Warning: Value out of range
is displayed.
Calibration successful
If the calibration has succeeded, you have two options:
-
Accept the new offset value. Select
Save
and go to the step 5/5.
The new offset value is displayed.
The date of the last calibration is updated. See chpt.
-
Reject the new offset value. Select
Cancel
and do a new calibration.