25
pH sensor - parameter
→
Select
Activation flags
. Select
High ref. impedance
. Validate with
.
Do not configure and activate the low error limit.
4.
Configure and activate the high error limit of the impedance value.
→
Go to device
pH sensor MS01
function
pH sensor
tab
Parameter
.
→
Select
Error triggers
.
→
Select
High ref. impedance
and enter the impedance value, above which an error event is generated.
Validate with
.
→
Select
Activation flags
. Select
High ref. impedance
. Validate with
.
5.
Enable the diagnostics
. Refer to chpt.
When the impedance value is more than one of the set limits, a warning event or an error event is gen-
erated. Details about the warning event and error event are given in chpt.
10.4
Monitor the pH value of the water sample
Do the following procedure:
1. Make sure that the login user level is
Installer
. Refer to chpt.
2.
Configure and activate the warning limits of the pH value.
→
Go to device
pH sensor MS01
function
pH sensor
tab
Parameter
.
→
Select
Warning triggers
.
→
Select
Low pH
and enter a pH value, under which a warning event is generated. Validate with
.
→
Select
High pH
and enter a pH value, above which a warning event is generated. Validate with
.
→
Select
Activation flags
. Select
Low pH
and
High pH
. Validate with
.
When the pH value is less than the set limit, a warning event is generated. Details about the warning
event are given in chpt.
If the pH value is more than the set limit, then a warning event is generated. Details about the warning
event are given in chpt.
English
Type MS01