Chapter 7 | Repair
Tests
Electrical Insulation Test
1. Connect the ViroVac
©
Surgical Smoke Evacuator
with a hipot tester according to the tester’s
instructions.
2. Turn on the ViroVac
©
Surgical Smoke Evacuator
.
3. Verify that the dielectric withstand is 1.5 kV at 1 second.
www.buffalofilter.com +1 716.835.7000 (International) 800.343.2324 (United States) [email protected]
Leakage Current Test
1. Connect the ViroVac
©
Surgical Smoke Evacuator
with an electrical safety tester according to the
tester’s instructions.
2. Turn on the ViroVac
©
Surgical Smoke Evacuator
.
3. Verify that the device leakage current is
≤
3 mA.
7-3