Chapter 3 | System Information
www.buffalofilter.com +1 716.835.7000 (International) 800.343.2324 (United States) [email protected]
System References
(continued)
Electromagnetic Compatibility
(continued)
Table 3
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The ViroVac
©
Surgical Smoke Evacuator
is intended for use in the electromagnetic environment specified
below. The customer or user of the ViroVac
©
Surgical Smoke Evacuator
should ensure that it is used in
such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment - Guidance
Portable and mobile RF communications
equipment should be used no closer to any part of
the
ViroVac
©
Surgical Smoke Evacuator
including
cables, than the recommended separation
distance calculated from the equation applicable
to the frequency of the transmitter.
Radiated RF
3 V/m
d = 1.7
√
P 80 MHz to 800 MHz
IEC 61000-4-3 80 MHz to 2.5 GHz 3 V/m
d = 2.3
√
P 800 MHz to 2.5 GHz
3 Vrms
d = [3.5/V1}
√
P
Conducted RF
Where P is the maximum output power rating
of the transmitter in Watts (W) according to
the transmitter manufacturer, and D is the
recommended separation distance in meters (M).
IEC 61000-4-6
150 kHz to 80 MHz
Field strength from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Field strengths from fixed transmitters, such as base stations for radio cellular / cordless telephones
and land mobile radios, amateur radios, AM and FM radio broadcasts, and TV broadcasts cannot be
predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
ViroVac
©
Surgical Smoke Evacuator
is used exceeds the applicable RF compliance level above, the ViroVac
©
Surgical Smoke Evacuator
should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the ViroVac
©
Surgical Smoke Evacuator
.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
NOTE:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE:
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structure, objects, and people.
3-6