Chapter 7 | Repair
www.buffalofilter.com +1 716.835.7000 (International) 800.343.2324 (United States) [email protected]
7-1
7
Repair
Precautionary Warnings
NOTE:
Proper care and maintenance of the ViroVac
©
Surgical Smoke Evacuator
is required
to ensure safe operation and service. Function and operation of the ViroVac
©
Surgical Smoke
Evacuator
shall be checked after each replacement session to protect patient and user.
Always switch off and unplug the ViroVac
©
Surgical Smoke Evacuator
before
disassembling or reassembling.
Wait at least three (3) minutes to allow stored energy to dissipate after the power is
disconnected before replacing any parts.
Do not touch any exposed wiring or conductive surfaces while ViroVac
©
Surgical Smoke
Evacuator
is disassembled and energized. Never wear a grounding strap when working
on an energized unit.
Perform a hipot test and current leakage test if the following parts have been replaced:
• PCB
• Motor