Help with faults | 9
Operation Manual UniversalExtractor E-800
77/88
Problem
Possible cause
Action
Boiling retardation
Incorrect positioning of the
beaker
u
Place the beaker directly
onto the heating plate.
Too long waiting time between
preparation of the instrument
and start
u
Start the instrument quickly
after solvent was added.
If dichloromethane is used
u
Use boiling aids, e.g. PTFE
boiling chips.
Dirty boiling surface
u
Replace beaker.
u
Use boiling aids, e.g. PTFE
boiling chips.
Worn boiling surface
No boiling
Incorrect positioning of the
beaker
u
Place the beaker directly
onto the heating plate.
Beaker position is not active
u
Activate the beaker position.
Heater power is set too low
u
Adapt the heating power to
the ambient conditions
(temperature, level above
sea).
u
Choose correct solvent.
9.2
Draining solvent from extraction glass chamber universal
Navigation path
➔
➔
[Service]
➔
[Actors]
➔
[Valves]
Precondition:
R
Solvent remains in the extraction glass chamber.
R
The lift is in the lower position.
R
The instrument has ambient temperature.
R
A beaker is placed under the extraction glass chamber Soxhlet position.
u
Navigate to
[Valves]
via the navigation path.
ð
The display shows the view
Valves
.
u
Select the position of the Soxhlet valve you wish to open.
u
Open the Soxhlet valve.
u
Wait until the extraction glass chamber is empty.
u
Close Soxhlet valve.