5 | Installation
Büchi Labortechnik AG
32/88
Operation Manual UniversalExtractor E-800
Precondition:
R
The coolant connection complies with the
specified parameters. See Chapter 3.5
"Technical data", page 25
R
Make sure that the instrument is not
connected to the power supply.
u
Install the inlet tube onto the connection
marked CW IN.
u
Attach the inlet tube in place with a hose clip.
u
Install the drain tube onto the connection
marked CW OUT.
u
Attach the drain tube in place with a hose clip.
5.6
Connecting the solvent tubes to the condenser tank bottle
NOTE
Leave the connection nuts in place when disassembling.
Precondition:
R
Make sure that the instrument is not
connected to the power supply.
u
Attach the inlet tube onto the connection
marked TANK IN.
u
Attach the drain tube onto the connection
marked TANK OUT (SOLVENT).
u
Attach a solvent connection nut with seals to
the connection marked OUT.
u
Attach a solvent connection nut with seals to
the connection marked IN (SOLVENT) .
u
Attach the inlet tube onto the connection
marked OUT .
u
Attach the drain tube onto the connection
marked IN (SOLVENT) .