BTL - 09
user manual & user's guide
page 9 of 42
II. APPLICATORS
INSTRUCTIONS FOR USE
•
Connect / disconnect applicators only when the unit is switched off.
•
Avoid direct contact between patient and applicator.
•
Do not strain mechanically movable inlet cable and connector.
•
Protect applicator’s surface against mechanical damage.
•
Always check applicator, movable cable and connector before use. Do not use applicator in case of any damage
to prevent injury with electrical current.
•
If adapter plug is not used, "Intensity" (12) display shows the highest value of magnetic induction intensity (mT).
•
If adapter plug is used, "Intensity" (12) display shows the intensity of magnetic induction in %.
•
Pulse magnetic field can cause disturbances, and in extreme case damage, to electronic instruments exposed to
it.
•
Check applicator’s temperature after termination of therapy. If the temperature is around 40°C or more, let the
applicator cool down before another use.
•
Have the applicator checked every 6 months by a qualified service person.
TECHNICAL DATA
Field outside application area
The cylinder coils, i.e. both solenoids and ring, do not have their corresponding levels of magnetic induction stated
outside the coils due to the fact that nearly all intensity of magnetic field is concentrated into the inner area of coil.
A significant value of magnetic induction outside coil can be also found very close to the coil´s surface.
Since flat coils´ magnetic field is symmetrical according the middle plane of coil, the diagrams of magnetic induction
levels show just one half of it. The other half is symmetrical. Generally speaking, value of magnetic induction drops
with the square of distance from coil. When it is clear from a diagram that the value of 2 mT is in the l10 cm distance,
then in the distance of 100 cm the magnetic induction is (10/100)
2
= 0.02 mT. This value has, due to short application
period, neglecting biological effects.
Used material
Upholstering material used for applicators is high quality inflammable artificial leather. All material used is harmless to
human health.