BTL - 09
user manual & user's guide
page 16 of 42
TRIPLE COVER (6)
Dimensions: 590x280x20
mm
Weight: 5.5
kg
Inductance: 51
mH
Resistance: 6.6
Ω
Max. voltage:
80 V
Max. current:
6 A
Max. power:
45 VA
Max. temperature:
45 C
Insulation: class
II
Environment: standard
Three independent elliptical coils with additional insulation from both sides are built-in to an upholstered, from both
sides insulated, mat. This insulation makes, together with internal parts, relatively rigid structure, which is still partially
flexible.
The coils are poled alternatively. Fig. 1 shows setting of magnetic induction levels valid for all intensity values.
The numeric values are relevant to the maximum actuation of the applicator. The higher values are valid for use
without adapter plug, the lower values are valid for use with adapter plug.
Electrical current induced in tissues is in direct relation to speed of change of magnetic induction shown at fig. 2.
ma
x.
in
te
ns
ity
m
ax
. i
nt
ensi
ty i
f
adap
te
r
is
u
se
d
Fig. 1: Triple cover
Fig. 2: Triple cover