v3.20z8JD10/03/2005EN
user manual
&
user's guide
Page 1: ...v3 20z8JD10 03 2005EN user manual user s guide...
Page 2: ...ns 7 Warranty 8 II APPLICATORS 9 Instructions for use 9 Technical data 9 Field outside application area 9 Used material 9 Big solenoid 1 10 Small solenoid 2 11 Ring 3 12 Disc 4 13 Double cover 5 14 Tr...
Page 3: ...tput 15 Control lights to signal output failure OUTWARD APPEARANCE BTL 09 is built in a lacquered metal box Signals displays controls and connectors for applicators or adapter plug are located on the...
Page 4: ...ven later in this manual Code Description 1 Big solenoid 50 cm diam 2 Small solenoid 30 cm diam 3 Ring 30 cm diam 4 Disc 24x24 cm 5 Double cover 2x24x17 cm 6 Triple cover 59x28 cm 7 Adapter plug Plug...
Page 5: ...ime has elapsed INTENSITY SETTING Intensity of magnetic field is shown on Intensity display 12 Intensity displayed is the maximum intensity at applicator s surface Intensity around individual applicat...
Page 6: ...push button there may not be any therapy going on and the control light Channel 6 of relevant channel must be on To switch to another channel press the push button 5 SHORT INSTRUCTIONS FOR USE 1 Conn...
Page 7: ...defect remains only at one channel you can use the second channel as an emergency No applicator may be connected to the faulty outlet TECHNICAL SPECIFICATIONS Number of independent outputs 2 Magnetic...
Page 8: ...or fax the fully completed warranty registration form within 30 days of installation All costs of shipment are the responsibility of the purchaser Damage to any part such as by accident or misuse or i...
Page 9: ...use Have the applicator checked every 6 months by a qualified service person TECHNICAL DATA Field outside application area The cylinder coils i e both solenoids and ring do not have their correspondi...
Page 10: ...of hard polystyrene that forms additional insulation Internal part of applicator is upholstered Movable outlet is provided by rubber outlet Fig 1 shows levels of magnetic induction valid for all inte...
Page 11: ...f hard polystyrene that forms additional insulation Internal part of applicator is upholstered Movable outlet is provided by rubber outlet Fig 1 shows levels of magnetic induction valid for all intens...
Page 12: ...f supporting coil with double insulation Fig 1 shows setting of levels of magnetic induction valid for all intensity values The numbers relate to maximum actuation of applicator The higher values rela...
Page 13: ...flat coil built in to both side upholstered pillow Fig 1 shows setting of levels of magnetic induction valid for all intensity values The numbers relate to maximum actuation of applicator The higher...
Page 14: ...magnetic induction levels valid for all intensity values The numeric values relate to the maximum actuation of applicator The higher values relate to use without adapter plug the lower values relate...
Page 15: ...BTL 09 user manual user s guide page 15 of 42 max intensity if adapter plug is used max intensity Fig 1c Double cover Fig 2 Double cover...
Page 16: ...makes together with internal parts relatively rigid structure which is still partially flexible The coils are poled alternatively Fig 1 shows setting of magnetic induction levels valid for all intens...
Page 17: ...immunity system or changes in fagocytosis etc Electromagnetic waving means the spreading of a variable magnetic field through space and it carries energy The irradiated tissue undergoes a number of ph...
Page 18: ...netic field improves information transfer of nerve stimulation muscle ability to contract blood circulation erythropoiesis stimulation improves tissue pH improves the immunity system through hypophysi...
Page 19: ...very suitable to combine with an intensive rehabilitation water treatment acupuncture physiotherapy etc The combined use of antibiotics and the pulse magnetic field has proved very efficient for it l...
Page 20: ...not to be shortened all complications will not necessarily disappear immediately Instead symptoms may fade away gradually after therapy has been completed After the healing process has been triggered...
Page 21: ...aena 26 Morbus Perthes 27 Myalgia 28 Myopathia 29 Neurodermatitis 30 Osteochondrosis 31 Osteomyelitis acuta chronica 32 Osteonecrosis aseptica 33 Osteoplastica 34 Osteoporosis 35 Osteosynthesis 36 Ost...
Page 22: ...here is a clinical finding but no morphological change has yet occurred apply magnetotherapy to the entire body in combination with non steroid anti rheumatoid medicaments such as Voltaren or Diclofen...
Page 23: ...15 Duration 30 45 minutes a day Application max 9mT for 45 minutes or max 11mT for 40 minutes Apply according to a location mostly big solenoid max 13mT for 20 minutes 11 Degeneratio spinalis Program...
Page 24: ...actura fractures in casts or with an osteotomy Program 15 or 19 Duration 30 max 40 minutes per day Therapy can be repeated Application max 11mT program 15 max 7mT program 19 for UL and LL small soleno...
Page 25: ...ion 9 12 mT for applicators other than solenoids 4 6 mT for solenoids Place the applicator mainly in the spinal area using a solenoid or double cover for feeble patients 29 Neurodermatitis Program 14...
Page 26: ...ay For example 3 times a day max for 90 minutes per session Application 3 5mT Comment It is always necessary to remove sequesters and evacuate hollows with a pus content Plaster or metal implants have...
Page 27: ...lly improves mood and imparts a sense of well being However expect a temporary worsening after 2 or 3 applications Do not change the program or cease treatment but intensity can be lowered 50 and retu...
Page 28: ...am 24 max 18mT for 30 minutes 53 Sy vertebralis painful vertebral syndrome with muscle spasms with nerve root stimulation Program 8 or 30 Duration 30 minutes Application triple cover program 8 max 14m...
Page 29: ...utes for ring Comment Very effective in early stages to decrease inflammation of the mucous membranes 60 Periostitis Program 1 Duration 5 8 therapy sessions for 30 minutes Application 6mT Use of Appli...
Page 30: ...lses 20ms with 20ms break frequency 25Hz Program 2 20x pulse 10ms with 10ms break break 100ms Program 3 10x pulse 20ms with 20ms break 150ms break Program 4 60ms pulse with 60ms break 5x pulse 10ms wi...
Page 31: ...pulse with 100ms break Program 8 30ms pulse with 60ms break frequency 11 1Hz Program 9 3x 20ms pulse with 20ms break 60ms break 10x 5ms pulse with 5ms break 120ms break Program 10 10ms pulse with 20ms...
Page 32: ...equency 3 5Hz 4x 71ms pulse with 71ms break frequency 7Hz 8x 35ms pulse with 36ms break frequency 14Hz 13x 5ms pulse with 30ms break frequency 28 5Hz Program 14 10ms pulse with 10 23ms break with repe...
Page 33: ...ith 212ms break frequency 4 6Hz Program 18 3x 35ms pulse with 115 150 break 15x 5ms pulse with 145ms break frequency 6 6Hz Program 19 15ms pulse with 40ms break frequency 18 2Hz Program 20 3x 65ms pul...
Page 34: ...4x 35ms pulse with 35ms break 35ms pulse with 80ms break 4x 18ms pulse with 18ms break 18ms pulse with 500ms break Program 24 10x 9ms pulse with 9ms break 100ms break 40ms pulse with 80ms break 2x 40...
Page 35: ...equency 6 6Hz Program 28 15ms pulse with 28ms break frequency 23 25Hz Program 29 5x 5ms pulse with 5ms break frequency 100Hz 80ms Program 30 20ms pulse with 200ms break frequency 4 5Hz Program 31 10x...
Page 36: ...frequency 25Hz 406ms 3x 16ms pulse with 510ms break frequency 1 9Hz Program 35 3x 20ms pulse with 20ms break frequency 25Hz 183ms break 4x 5ms pulse with 298ms break frequency 3 3Hz Program 36 5x 6ms...
Page 37: ...uency 23 3Hz 189x 7ms pulse with 30ms break frequency 27Hz 220x 15ms pulse with 15ms break frequency 33 3Hz Program 41 10x 5ms pulse with 5ms break 212ms break 10x 10ms pulse with 10ms break 13x 5ms p...
Page 38: ...17ms pulse with 44ms break frequency 16 4Hz Program 45 5x 6ms pulse with 4ms break 20ms break Program 46 7x 30ms pulse with 20ms break 217ms break 12x 5ms pulse with 212ms break Program 47 5x 40ms pul...
Page 39: ...476ms break 3x 5ms pulse with 471ms break frequency 2 1Hz Program 52 10x 5ms pulse with 10ms break frequency 66 6Hz 100ms 60ms pulse with 80ms break 60ms pulse with 100ms break 60ms pulse with 80ms b...
Page 40: ...with 295ms break frequency 3 3Hz Program 56 10ms pulse with 155ms break frequency 6 1Hz Program 57 5x 20ms pulse with 100ms break frequency 8 3Hz 1065ms break Program 58 5x 7ms pulse with 4ms break 9...
Page 41: ...th 15ms break frequency 50Hz Program 63 100x 5ms pulse with 35ms break frequency 25Hz 100x 5ms pulse with 5ms break frequency 100Hz Program 64 5ms pulse with 20 45 ms break frequency 20 40Hz 120s peri...
Page 42: ...any means incl electronic mechanic photographic or other records without previous approval from BTL Industries Limited BTL Industries Limited operates a policy of continuous development Therefore it r...