page 7 of 42
TROUBLE SHOOTING
This section will try to anticipate potential problems that you may encounter in the day-to-day use of the product.
Included is information which should help you solve these problems. If after trying suggested solutions, the product still
does not function properly, please contact your service center.
Symptom:
Solution:
"Intensity" display (12) shows "--".
Check connection of applicators; connect different
applicator.
Red control light "Error A" or "Error B" (15) is on and
"Intensity" display (12) does not show "Er".
Connect both applicators to adapter plug; connect
different applicator.
Power switch is on, front panel is dark.
Check the power connection to the instrument and to
the mains socket; check the voltage in socket; check
fuses.
Red control light "Error A" or "Error B" (15) is on,
"Intensity" display (12) shows "Er" and "Time / WS"
display (8) does not show "05".
Check connection of applicators; connect different
applicator.
Red control lights "Error A" or "Error B" (15) are on,
"Intensity" display (12) shows "Er" and "Time / WS"
display (8) does not show "05".
Switch the instrument off and on again after around 5
sec. If defect remains only at one channel, you can use
the second channel as an emergency. No applicator
may be connected to the faulty outlet!
TECHNICAL SPECIFICATIONS
Number of independent outputs:
2
Magnetic induction max:
(depends on the type of applicator and type of current)
Time of therapy max:
5 to 90 min. (depends on the type of applicator and type of
current)
Accuracy of time therapy:
±
3 %
Magnetic field frequency:
1 Hz to 60 Hz
Accuracy of generated frequency:
±
3 %
Mains voltage:
198 V to 252 V (complies with EN)
50 Hz
±
10 %
Power input:
350 VA typical
650 VA max
Operating environmental conditions:
temperature:
+ 15
°
C to + 40
°
C
relative humidity:
30 % to 75 %
atmospheric pressure:
700 hPa to 1060 hPa
Storage/transport environmental conditions:
temperature:
- 10
°
C to + 55
°
C
relative humidity:
5 % to 85 %
atmospheric pressure:
650 hPa to 1100 hPa
Dimensions:
375x110x225 mm
Weight:
8.5 kg approx.
Noise level:
54 dB max (measured in the acoustic tile room 300 m
3
min)
The unit is designed to comply with:
IEC 601-1 (1994), IEC 601-1 +A1, A11, A12 (1995), CISPR 14
Equipment Class:
Class I
Type B applied part:
BF (symbol
) - This parameter is guaranteed only with
original BTL applicators)
Protection:
IP 20
Mains transformer fuses used:
primary side (external):
2x T6.3A for 230 V / 2x T10A for 115V
ON indication (green colour):
LED indicator POWER
In-use indication:
A and B channel bar-graphs