BTL - 09
user manual & user's guide
page 6 of 42
The table below shows the transfer of set intensity of magnetic induction (in %) to mT value for an individual
applicator:
Intensity set
[%]
Big solenoid
[mT]
Small solenoid
[mT]
Ring
[mT]
Single cover
[mT]
Double cover
[mT]
Triple cover
[mT]
0 0.0 0.0 0.0 0.0 0.0 0.0
10 0.6 0.8 1.2 1.3 1.2 1.2
20 1.2 1.6 2.4 2.6 2.4 2.4
30 1.8 2.4 3.6 3.9 3.6 3.6
40 2.4 3.2 4.8 5.2 4.8 4.8
50 3.0 4.0 6.0 6.5 6.0 6.0
60 3.6 4.8 7.2 7.8 7.2 7.2
70 4.2 5.6 8.4 9.1 8.4 8.4
80 4.8 6.4 9.6 10.4 9.6 9.6
90 5.4 7.2 10.8 11.7 10.8 10.8
100 6.0 8.0 12.0 13.0 12.0 12.0
TO START AND INTERRUPT THERAPY
Start the therapy by pressing the "Start/Pause" push-button (11). The bar-graph (14) is blinking during therapy.
You cannot change parameters during therapy. The therapy ends automatically when the pre-set time has elapsed.
Press "Start/Pause" push-button (11) repeatedly to interrupt the therapy at any time. You may change the parameters
or applicators during an interruption. Press "Start/Pause" push-button (11) again to continue therapy.
If an applicator is completely covered (e.g. by patient’s body), it can become -- particularly during higher intensities --
overheated (above 40°C). This is signaled by yellow control light on top of bar-graph (14). Disconnect the applicator
and let it cool down.
TO RESET VALUES OF THE LAST THERAPY
Press the push-button with arrow (13) to set values of the last therapy. To use the push-button, there may not be any
therapy going on, and the control light "Channel„ (6) of relevant channel must be on. To switch to another channel,
press the push-button (5).
SHORT INSTRUCTIONS FOR USE
1. Connect the instrument to mains.
2. Set the power switch into position 1.
3. Connect
applicator(s).
4. Press push-button (5) to select between channels "A" and „B".
5. Press push-button "P/F"(2) to select "Program" or "Frequency" mode.
6. Press buttons (3) to select required program or frequency.
7. Set therapy time by push-buttons (7).
8. Press push-buttons (9) and (7) to set value of variable part of frequency.
9. Set intensity of magnetic field by push-button (10).
10. Start therapy with "Start/Pause" push-button (11).
11. Interrupt therapy at any time by "Start/Pause" push-button (11).
12. Press push-button (13) to reset values of the last therapy.
FUSE REPLACEMENT
Fuses are located on the rear panel in black round cases. Make sure that power switch is set to “0“. Unplug the mains
cable from the instrument and from the mains plug. Use an appropriate screwdriver to rotate the inner segment of the
fuse case anti clockwise to remove and replace the fuse. Do not use fuses other than those stated in the label below
the fuse cases. Only qualified personnel should do this procedure.
MAINTENANCE AND TRANSPORT
The unit may be cleaned by wiping over with a damp cloth. Do not use alcohol-based solutions. Clean applicators
when disconnected from instrument. Use sponge with a cleaning agent. Disinfect applicators using 10% solution
of e.g., Persteril.
Transport the instrument in original packaging. Remove power cable and aplicators. Avoid rough handling.