Installation Instructions
22
User Manual CPM II2066en Rev.B
© BTG 2019
2.5.4
Sensor Control Module SCM-8000
NOTE!
These connection instructions are only valid for DRT-5500. Other instruments
equipped with the SCM-8000 have all connections to the SCM unit pre-wired
at delivery.
NOTE!
Before installation, ensure that all power to the system has been turned off.
Cable connections must be made by authorized personnel.
DANGER
Hazardous voltage
in the equipment
HCM
8000
DI3
DO
+ -
+ -
+ -
+ -
+ -
+ -
DI1 DI2
AO
AI
P1
P2
P3
RS-485
red
black
blue
white
+ -
+ -
+ -
+ -
+ -
+ -
P1
P2
P3
+
-
-
-
+
-
-
-
+
-
-
-
P6
P5
P4
SCM-8000
Common
Solenoid
valve
2
Solenoid
valve
3
Solenoid
valve
5
Solenoid
valve
6
Solenoid
valve
7
Solenoid
valve
4
Solenoid
valve
9
Solenoid
valve
8
Solenoid
valve
1
-
+
-
P4
-
+
-
P6
-
+
-
P5
-
-
-
Sensor
Cable
SCM-8000
-
+
-
+
P3
Sample
button
Sampling
indicatior
+24
V
-
+
-
+
P2
-
+
-
+
P1
brown
black
white
blue
H
CM
-
8000
P1
P2
P3
P4
P5
P6
S
CM-8
000
SCM
8000
Fig 19 Connection of
SCM-8000