Product Introduction
© BTG 2019
User Manual CPM PI2066en Rev.B
5
Support System Specifications
Supply Voltage
Power supply unit 100 - 240 V AC, 50-60 Hz.
AC input range: 90 - 264 V continuous operation.
Disconnecting Device
An external 2-pole switch close to the CPM is required. The switch must be approved in
accordance with the IEC 60947-2 and IEC 60947-3 requirements.
Power Consumption
Maximum: 135 VA (CPM-1300), 515 VA (CPM-1400), 160 VA (CPM-1500)
Altitude
0 to 2000 m (0-6560ft) without any restrictions.
2000 to 6000 m (6560 to 20000ft) reduce output power or ambient temperature.
Altitude de-rating = 5 W / 1000 m or 5 °C / 1000 m.
Humidity
5 to 95% r.h (IEC 60068-2-30)
Over-voltage Category
Category III: IEC 62103, EN 50178, altitudes up to 2000 m
Category II: altitudes from 2000 m to 6000 m
Degree of Pollution
2: IEC 62103, EN 50178, not conductive
Physical Specifications
Materials
Casing: Polycarbonate thermo plastic
Cable fittings: Polyamide thermo plastic
Storage Temperature
Max. 80 °C (176 °F)
Min. -25 °C (-13 °F)
Operation Temperature
Max. 50 °C (122 °F)
Min. 0 °C (32 °F)
Degree of Protection
IP 65, comparable to NEMA 4x and better, the CPM is intended for use indoors.
Weight
CPM: 2 - 2.5 kg (4.4 - 5.5 lbs) depending on configuration
Cables
Power supply flexible cable: 0.3 - 2.5 mm
2
(AWG = 28-12)
Signalling cable: 0.2 - 2.5 mm
2
(AWG = 24-12)
Transmitter Cable
CPM-1300: Standard length: 10 m [33 ft] (option up to 55 m [180 ft])
Cable diameter: 5.5 mm [0.22"]
Connector diameter: 15 mm [0.6"]
CPM-1400: Standard length: 10 m [33 ft] (option up to 20 m [66 ft])
Cable diameter: 13 mm [0.5"]
No fixed connector
CPM-1500: Standard length: 10 m [33 ft] (option up to 20 m [66 ft])
Cable diameter: 5.5 mm [0.22"]
No fixed connector
Cable Inlets
There are cable glands for signal cables (diameter 4-8 mm) and for power supply cable
(diameter 4-12 mm) in the bottom of the CPM.