HRB Bench Scale Installation Instructions
11
3.1 Base Communications setting
The HRB base has to be linked to an approved high end indicator. It uses one of the
dedicated RS232 channels to correctly communicate to the base.
Use one of the dedicated RS232 base to indicator cables as listed below
AWT25-501983
RS232 Base to indicator cable (keyed RJ45 to striped end) 10ft
AWT25-502071
RS232 Base to indicator cable (keyed RJ45 to striped end) 25ft
AWT25-502072
RS232 Base to indicator cable (keyed RJ45 to striped end) 50ft
These cables have the correct keyed RJ45 connector. Pictured below are the cable
ends ready for connection to the HRB and indicator.
3.1 Base Communications setting
Baud rate:
115200
Data bits:
8
Party:
None
Stop Bits:
1
3.2 Base configuration
All base configuration will be found within the indicator, please refer to the indicator
service manual or HRB to indicator tech guides for base setup and calibration.
BROWN
WHITE/BROWN
GREEN
WHITE/BLUE
BLUE
WHITE/GREEN
ORANGE
WHITE/ORANGE
Keyed RJ45 Connector
Key
Summary of Contents for HRB