157
Bruker Optik GmbH
HYPERION User Manual
Troubleshooting 7
7.2.7
Problems regarding mapping measurements
Mapping measurements without ATR objective
Mapping measurements with ATR objective
Possible cause
Solution
After a PC reboot, the motorized stage has not
been reset.
Reset the motorized stage.
☞
See section 4.5.
Note:
The motorized stage needs to be reset
after each PC reboot!
In OPUS, an experiment file is loaded of which
the parameter settings are not defined for map-
ping measurements (i.e. the selected
Imaging
Device
option is not for mapping measure-
ments).
In OPUS, load an experiment file of which the
parameter settings are defined for a mapping
measurement.
☞
See appendix B.
Possible cause
Solution
After a PC reboot, the motorized stage has not
been reset.
Reset the motorized stage.
☞
See section 4.5.
Note:
The motorized stage needs to be reset
after each PC reboot!
At the mapping measurement start, the dis-
tance between sample and crystal tip (with the
ATR crystal being in the measuring mode posi-
tion) was more than 2 mm.
Abort the measurement. Move the stage
upwards until the distance between sample
and crystal tip is
≤
2 mm. Restart the mapping
measurement.
Hardware Setup in OPUS: The wrong stage
has been selected in the setup of the
Mapping
Device
and
Imaging device
.
OPUS: Both in the mapping device setup and
in the imaging device setup, select the stage
option
LStepATRStage
(and not
LStepStage
!).
☞
For information about how to setup the
mapping device and the imaging device,
refer to the OPUS/VIDEO Manual.
In OPUS, an experiment file is loaded of which
the selected imaging device is not defined for
mapping measurements with ATR objective.
OPUS: Select an imaging device which is
defined for mapping measurements with ATR
objective.
Hardware Setup in OPUS: The option
Autofo-
cus
has not been selected in the setup of the
Imaging device
.
Note: Autofocus is an electronic option.
Select an imaging device for which this elec-
tronic option is selected.
If such an imaging device does not yet exist,
set it up.
☞
For information about how to setup the
imaging device, refer to the OPUS/VIDEO
Manual.
Summary of Contents for HYPERION
Page 1: ...HYPERION User Manual I 24319 ...
Page 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Page 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Page 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Page 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Page 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Page 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Page 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...