21
Bruker Optik GmbH
HYPERION User Manual
Installation 2
1
Connect the
signal cable
of the FPA
detector to the corresponding socket at
the PC.
2
Connect the
trigger cable
to the
RS42
Sync Input
socket
②
at the bottom side
of the FPA detector compartment as well
as to the corresponding socket
③
at the
spectrometer rear side.
➣
Attention:
Before connecting the
trigger cable, make sure that the
spectrometer is switched off!
Note: In case of a spectrometer of the
TENSOR series, a distributor box is
required.
3
Connect the
power supply cable
to the
Power
socket
④
at the bottom side of
the FPA detector compartment.
The power is supplied by a dedicated
external power supply unit. Connect this
external power supply unit to a mains
socket outlet.
Important note:
Connect the power sup-
ply cable only to a socket outlet with
earthing contact
⑤
.
Make sure that the
socket complies with IEC.
①
③
②
④
⑤