HYPERION User Manual
Bruker Optik GmbH
Measurement parameters B
Modifying an experiment file
Normally, the default parameter settings (especially the optics parameter) of the deliv-
ered experiment files need to be adapted to configuration of the analysis system in ques-
tion. To do this, proceed as follows:
1.
Select in the OPUS
Measure
menu the
Video-assisted measurement
function
.
2.
Load a supplied experiment which is appropriate for the measurement mode and
the measurement technique (transmission, reflection, mapping measurement
1
,
measurement with ATR objective, measurement with FPA detector
2
) you intend to
perform.
3.
Select the corresponding parameter options. (See the following sections.)
4.
Save the modified parameter settings. To do this, click in the
Video assisted Mea-
surement
dialog on the
Advanced
tab. On this dialog page, click on the
Save
button.
(Note: It is advisable to save the modified experiment file with a different file name.)
Information about the optics parameter
To modify the optics parameter, click on the
Optics
tab. See the following figure.
1. Mapping measurements are only possible if the microscope is equipped with a motorized sample stage.
2. The FPA detector is only available for the microscope variant HYPERION 3000.
Figure A.1: OPUS dialog
Video assisted Measurements
- Dialog page
Optics parameter
Summary of Contents for HYPERION
Page 1: ...HYPERION User Manual I 24319 ...
Page 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Page 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Page 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Page 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Page 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Page 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Page 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...