9
Bruker Optik GmbH
HYPERION User Manual
2
Installation
2.1
General information
The installation of the complete analysis system (i.e. microscope plus FT IR spectrome-
ter plus data station plus possible accessories) as well as the initial start-up and the site
acceptance test are done by the Bruker service. The operating company has to provide
an installation site that meets the site requirements described in section 2.5.
☞
See also the technical document
Installation Requirements for IR-Microscope
HYPERION 1000 / 2000 / 3000
provided by Bruker Optik GmbH in advance of the
instrument delivery.
The installation of the microscope includes the following actions:
• connecting the microscope to the power supply
• connecting the microscope to the purge gas supply line (Not required, but highly
recommended.)
• connecting the microscope to a computer
• connecting the microscope to the FT IR spectrometer
☞
For detailed information about how to connect the FT-IR spectrometer, the computer
and peripherals, refer to the corresponding manuals.
➣
Important note:
The operator can reconnect all cables if required, for example after
having relocated the analysis system. For information about the connections, refer
to the corresponding sections in this chapter.
2.2
Inspecting the packaging
After having received the microscope, inspect the packaging for damages. If there are
any signs of damage contact shipping company.
2.3
Transporting the microscope
CAUTION
Possible damage to the delivered microscope because of transport
damage
Non-observance of the following safety instructions could result in injury.
➣
A microscope delivered in a damaged packaging may be damaged as well.
Therefore, in this case do not put the microscope into operation. Contact Bruker
instead. For the contact data, see section 1.5.
CAUTION
Injury and/or microscope damage due to an inadequate method of
transport
Non-observance of the following safety instructions could result in injury.
➣
For a short-distance, the microscope can be carried by at least two persons.
➣
For a long-distance, put the microscope on a wheeled table, for example. To avoid
damages, transporting the microscope in original packing is recommended.
Summary of Contents for HYPERION
Page 1: ...HYPERION User Manual I 24319 ...
Page 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Page 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Page 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Page 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Page 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Page 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Page 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...