To activate the heater the following conditions have to be met:
1.
The magnet system must be deenergized. If the magnet is energized, the heater
cannot be activated, and the indicator remains inactive ("OFF").
2.
The magnet must be below ambient temperature (300 K). While the magnet is at
ambient temperature the heater and the indicator remain inactive. This is a
protection against overheating the magnet. It will also make sure that the heater is
switched off as soon as the magnet reaches ambient temperature (after 1 – 2 days).
Magnetic Field Indicator
The magnetic field indicator shows whenever the magnet is energized and hazardous
fringe fields occur around the magnet.
The indicator can be placed on top or on the front panel of the BCFC. Make sure that the
indicator is clearly visible from all locations within the 5 G contour (see Appendix
and
).
There is a connector for an additional indicator on the rear panel of the BCFC in case an
external indicator needs to be installed for enhanced visibility.
Emergency Quench and Status Indicators
In case of emergency (e.g. fire), it may be necessary to quench the magnet intentionally
and to eliminate the hazardous magnetic fringe field. The quench can be executed via
an emergency button ("Magnet Quench") on the front panel of the BCFC.
For remote control of the "Magnet Quench" option the BCFC is equipped with a
connector at the rear panel (see
).
Activating the "Magnet Quench" will lead to the following actions on the magnet system:
•
The compressor will be stopped immediately.
•
A heater is activated in the magnet system which rapidly increases the magnet
temperature. Raising the magnet’s temperature will lead to a quench of the magnet.
The heater is backed by an internal UPS in the BCFC, so the "Magnet Quench"
functionality is available even during power failure (see also section
).
Summary of Contents for EPR 12T Cryogen Free
Page 1: ...Version Innovation with Integrity EPR 12T Cryogen Free 01 User Manual EPR Spectroscopy ...
Page 6: ...6 ZTKS0213 Z31939 Rev 01 Contact ...
Page 25: ...25 ZTKS0213 Z31939 Rev 01 4 Assembling Approved persons Bruker Service only ...
Page 26: ...26 ZTKS0213 Z31939 Rev 01 Assembling ...
Page 48: ...48 ZTKS0213 Z31939 Rev 01 Operation ...
Page 54: ...54 ZTKS0213 Z31939 Rev 01 Maintenance ...
Page 55: ...55 ZTKS0213 Z31939 Rev 01 8 Disassembling Approved persons Bruker Service only ...
Page 56: ...56 ZTKS0213 Z31939 Rev 01 Disassembling ...
Page 58: ...58 ZTKS0213 Z31939 Rev 01 Warning Signs ...
Page 60: ...60 ZTKS0213 Z31939 Rev 01 List of Figures ...
Page 62: ...62 ZTKS0213 Z31939 Rev 01 List of Tables ...
Page 64: ...64 ZTKS0213 Z31939 Rev 01 Glossary ...
Page 80: ...Bruker Corporation info bruker com www bruker com Order No Z31939 ...