Figure 5.9:
Bruker Cryogen Free Controller (BCFC) – Rear View
Figure 5.10: Magnet quench indicators on the BCFC front panel
•
The green indicator (1, "ARMED") is active
while the "MAGNET QUENCH" button is in
stand by mode. It should always be active
("ON"), otherwise the magnet quench safety
functionality is not available.
•
The red indicator (2, "TRIGGERED") is
active when the "MAGNET QUENCH"
button has been pushed and the magnet is
going to quench.
•
The red indicator (3, "AC INPUT POWER
FAILURE") displays whether AC power is
supplied to the BCFC. As it has an internal
UPS, it will keep on working for some
residual time. After this time, the BCFC is
completely inactive! Restore AC power
quickly!
•
The yellow indicator (4, "REPLACE
BATTERY") lights up if the battery of the
UPS needs to be replaced (every 4 – 5
years).
NOTICE
Use the "MAGNET QUENCH" button only in case of emergency. It is a safety
element that is optimized for quick reaction time. The magnet system may get
damaged from this action.
Summary of Contents for EPR 12T Cryogen Free
Page 1: ...Version Innovation with Integrity EPR 12T Cryogen Free 01 User Manual EPR Spectroscopy ...
Page 6: ...6 ZTKS0213 Z31939 Rev 01 Contact ...
Page 25: ...25 ZTKS0213 Z31939 Rev 01 4 Assembling Approved persons Bruker Service only ...
Page 26: ...26 ZTKS0213 Z31939 Rev 01 Assembling ...
Page 48: ...48 ZTKS0213 Z31939 Rev 01 Operation ...
Page 54: ...54 ZTKS0213 Z31939 Rev 01 Maintenance ...
Page 55: ...55 ZTKS0213 Z31939 Rev 01 8 Disassembling Approved persons Bruker Service only ...
Page 56: ...56 ZTKS0213 Z31939 Rev 01 Disassembling ...
Page 58: ...58 ZTKS0213 Z31939 Rev 01 Warning Signs ...
Page 60: ...60 ZTKS0213 Z31939 Rev 01 List of Figures ...
Page 62: ...62 ZTKS0213 Z31939 Rev 01 List of Tables ...
Page 64: ...64 ZTKS0213 Z31939 Rev 01 Glossary ...
Page 80: ...Bruker Corporation info bruker com www bruker com Order No Z31939 ...