44
ZTKS0213 / Z31939 / Rev.: 01
Operation
5.6.5
Warming up the Magnet
Precondition
•
Magnet deenergized (refer to
"Energizing and Deenergizing in Main Mode" on
)
•
Magnet cold (below 300 K)
Procedure
1.
Turn off the compressor (see
).
2.
Activate the switch "WARM UP MAGNET". This switch is activated if the indicator
within the switch is active. If both preconditions are fulfilled, the heater is being
activated. Only exception: AC power failure.
It takes up to ~ 2 days for warming up the system. The heater is being deactivated if
the magnet is at 300 K.
3.
Deactivate the "WARM UP MAGNET" switch. Check if the indicator within the
switch is OFF:
Figure 5.18: After work / service: warming up the magnet system