32
ZTKS0213 / Z31939 / Rev.: 01
Operation
5.2.4
Removing the Vacuum Valve
Figure 5.6:
Removing the vacuum valve
1.
Vent the pumping line.
2.
Remove the pumping line at the KF 40
flange (7).
3.
Release the valve stem (6) from the sealing
plug (2) and pull out the valve stem.
4.
Remove the four M6 x 12 screws (4) from
the half rings (3).
5.
Remove the half rings (3) and remove the
vacuum valve.
Figure 5.7:
Mounting the earthing cable
6.
Mount the protective cap on the top plate.
7.
Mount the earthing cable.