5.6.4
Emergency Quench
Precondition
•
"HIGH MAGNETIC FIELD" indicator is ON;
•
Indicator "ARMED" is ON.
Procedure
1.
Push the magnet quench button.
2.
Check that the rhythmic noise of the compressor stops and the red indicator
("TRIGGERED") is ON.
3.
Wait until the magnetic field indicator is turned OFF. It can take up to a few minutes
until the magnetic fringe field has disappeared after pushing the emergency button
"MAGNET QUENCH".
Figure 5.17: Quenching the magnet in an emergency case