Installation Manual Version 001
BRUKER
129 (167)
13
Sample Heater
Option
13
General Information
13.1
The B-ACS SAMPLE HEATER 120C (
) is used in connection with the
BRUKER AUTOMATIC SAMPLE CHANGER (B-ACS 60, B-ACS 120) for NMR
Spectrometer‘s.
Using the B-ACS Sample Heater it is possible to preheat NMR probes to a range
of +30°C and +120°C depending upon the ambient temperature.
Prerequisites
13.1.1
Before installing the SAMPLE HEATER, the following points must be observed:
The B-ACS LIGHT BARRIER MAGAZINE.
The new Sample Changer must have a new version of the Light Barrier Magazine.
If this is not the case, it must be exchanged for our new one, because the Maga-
zine belt will not turn with the Sample Heater using the old Light Barrier.
Therefore if you are using the old version B-ACS 60 Part No. H1082, or the old
version B-ACS 120 Part No. H655, you will need to replace it with the new version
B-ACS 60 / Part No. H1402, or the B-ACS 120 Part No. H1403.
In addition, a sliding ring for transmission from the DC power supply, must be
placed directly under the cabinet (see
Summary of Contents for B-ACS 120
Page 1: ...BRUKER Version Bruker Automated Control Systems B ACS 60 120 Installation Manual 001...
Page 7: ...Installation Manual Version 001 BRUKER 7 167 1 Declaration of Conformity 1...
Page 10: ...10 167 BRUKER Installation Manual Version 001 Introduction...
Page 40: ...40 167 BRUKER Installation Manual Version 001 Operating Instructions...
Page 56: ...56 167 BRUKER Installation Manual Version 001 Special Tools...
Page 60: ...60 167 BRUKER Installation Manual Version 001 Technical Data...
Page 150: ...150 167 BRUKER Installation Manual Version 001...
Page 160: ...160 167 BRUKER Installation Manual Version 001 Figures...
Page 167: ...Installation Manual Version 001 BRUKER 167 167 Lastpage...