© Brüel & Kjær Vibro
● C107758.002 / V04
Page 23 of 42
Technical alterations reserved!
Brüel & Kjær Vibro
│
Installation instruction VCM-3 / VCM-3 Ex
Termination of sensor signals
EN
UNRESTRICTED DOCUMENT
7.5
Brüel & Kjær Vibro AS-247 dual axis accelerometers
on channel 11-12
AS-247 sensor mounting.
AS 247 is a bi-axial sensor, that is, it produces two output
signals one for each axis. In the figure these two axes are
denoted XSIG and ZSIG.
This sensor needs a +24 VDC supply which is available on
the (V+) terminals on AC/DC channel 11 and 12.
The color of the connection wires corresponds to sensor
connection cable AC-1401.
NOTE!
Maximum current output in this configuration
is 55 mA in total.