MANUAL P/N 316005 REV: B
13
1.3 Mounting Configurations
Figure 1-3 illustrates the various configurations available for mounting the 316 control unit and power supply.
Note that when installing a mounting tab (8) into the left side of the control unit, the door should be
open.
The bracket (9) used to connect two control units for side by side rack mounting is not symmetrical. Install as
shown.
The rear support bracket assembly consists of a U-shaped bracket (7), and two spring clamps (6) which grip
into grooves in the side of the control unit.
Note that the 316 Controller should be mounted in a location with free air flow and ambient
temperature less than 40
o
C.
1.4 Cable Connections
To avoid confusion in cable connections, a set of labels is provided with the cable. The labels are easily
attached to the gauge connector with the adhesive backing. The rear panels of the Convectron and
thermocouple modules are labeled with their respective display designations – A, B and C. See Figure 1-4 for
cable connections. Tighten the screws to secure the connectors.
Figure 1-4 Cable Connections