Section 4 Maintenance
Brooks
®
Ar-Mite
TM
MT3750C
28
Service Information
On Site Adjustment/Calibration
The transmitter can be adjusted and/or calibrated with the help of a Windows
95/98 calibration program in combination with an applicable interface that
can be plugged into any RS232 serial PC-port. The calibration data is stored
in the non-volatile memory in the transmitter. After calibration the transmitter
functions as a stand alone device
(consult factory for more details).
PC interface is needed to calibrate the transmitter. With this interface
the transmitter can be connected to the serial RS232 port of a computer.
At power/start-up the transmitter is waiting (~2 seconds) for serial
communication before entering normal operation. When a serial
communication is established, the transmitter is instructed to sustain its
low-power mode (typically 2.6 mA). In this low-power mode the transmitter is
directly powered from the RS232 port without the need of an external power
supply.
Software Kit (Calibration CD-ROM and RS232 Interface): partnumber:
F535Y001ZZZ