![Brooks 5851EM Series Installation And Operation Manual Download Page 30](http://html1.mh-extra.com/html/brooks/5851em-series/5851em-series_installation-and-operation-manual_2818708030.webp)
3-4
Brooks
®
Models 5965, 5851EM
Section 3 Operation
Installation and Operation Manual
X-TMF-5965-5851EM-MFC-eng
Part Number: 541B122AAG
September, 2009
Note 4: lf Brooks secondary electronics are being used as a power supply
during the calibration, the 5 volt reference must be enabled on the card
Edge version for proper operation (see Section 2-7). Remember to
deactivate the 5 Volt reference before installing the calibrated mass flow
controller in the system where Terminal 10 is grounded.
Figure 3-3 Card Edge PC Board Jumper Location and Function