3)
Raise the Vane Lid to its uppermost position by pressing the up arrow.
4)
Place the circular foam insert ( PFT-101 for 6” Trough and PFT-106 for 5” Trough)
into the Trough. Remove the Trough to accomplish this, if necessary, from the
PFT. Install the Trough on the PFT.
5)
Press the down arrow to lower the Vane Lid until it just makes contact with the
foam insert. Then press and hold the down arrow for 2 seconds. This will adjust
the position of the Vane Lid so that the foam is slightly compressed. Do not hold
longer than 2 seconds because this will cause further compression of the foam
insert, which is not desired.