EMC Information
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.
Emissions Test
Standard
Electromagnetic Environment- Guidance
RF emissions (radiated)
CISPR 11
EN55011
The device uses RF energy only for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic equipment.
RF emissions (conducted)
CISPR 11
EN55011
The device is suitable for use in all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network.
Harmonic emissions
IEC 61000-3-2
EN61000-3-2
Voltage fluctuations/Flicker
emissions
IEC 61000-3-3
EN61000-3-2
Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic Environment-
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input-output lines
±2 kV for supply mains
±1 kV for input/output lines
Mains power quality should be that of a typical home or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV for common mode
Mains power quality should be that of a typical home or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
>95% dip in Voltage for 0.5
periods
@ 230Vac and 100Vac
60% dip in Voltage for 5 periods
@ 230Vac and 100Vac
30% dip in Voltage for 25
periods
@ 230Vac and 100Vac
>95% dip in Voltage for 0.5
periods
@ 230Vac and 100Vac
60% dip in Voltage for 5 periods
@ 230Vac and 100Vac
30% dip in Voltage for 25
periods
@ 230Vac and 100Vac
Mains power quality should be that of a typical home or hospital
environment. If the user of the device requires continued operation
during power mains interruptions, it is recommended that the device
be powered from an uninterruptible power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical hospital or home environment.