27
External Battery
An external battery may be used to increase the running time. This battery will power
the NIPPY+ for the same time as an internal battery, depending on settings and leak.
External Battery, part number 0910.
External battery charger, part number 0911
These batteries should never be used to run any other type of equipment.
Instructions for Use
•
Connect the battery to the NIPPY+ Aux Power input.
•
Switch on the NIPPY
+. The Ext Batt light will flash and a “Running on battery
power” message will be displayed on the NIPPY+ screen. Press the alarm mute
button to hide the message.
•
To disconnect a battery:
Always switch off the ventilator first
. Press the plug
release button on the connector and withdraw the connector.
To Charge an External Battery
•
Connect the battery to the NIPPY+ whilst it is connected to mains electricity.
Alternatively, charge with the battery charger as follows:-
•
Place the charger on a smooth flat surface.
•
Connect the charger to the battery socket
before switching on the mains
power
.
•
Connect the mains plug to the AC supply and switch on.
•
Leave on charge until the charged / ready indicator lights.
•
Batteries may be left connected to the charger until required for use.
Safety
•
Warning! High voltages exist inside the charger.
•
Do not remove the cover. Return to Breas Medical Ltd if a fault occurs.
•
Do not expose to water or dust.
•
Do not cover the charger whilst in use
•
Do not use any other type of battery charger. This could lead to damage to the
battery and personal injury.
•
Do not attempt to connect any battery other than those supplied by the
manufacturer. Use of any other type of battery could lead to personal injury and
damage to the ventilator.
•
Ensure that the mains lead is not damaged.
•
Batteries may produce explosive gases during charging. Always charge away from
sparks or sources of ignition. Do not smoke near a battery whilst charging.
Disconnect from the mains before disconnecting the battery from the charger