17
Setting up the NIPPY ST+
in CPAP mode
1. Place the NIPPY ST+ on a clean, level surface.
2. Connect the breathing circuit tube to the outlet. It is recommended that a bacterial
filter be fitted between the outlet and the 22mm diameter breathing tube.
3. The breathing circuit must contain an intentional leak to prevent re-breathing of
exhaled air. This may be achieved with either an integrated leakage in the mask
or a leakage/exhalation port.
4.
Connect the mask to the outlet tubing on the breathing circuit.
Note: Do not fit the mask onto the patient at this point.
5. Press the Start/Stop button and select CPAP mode.
6.
Carry out alarm tests as described in “Alarm conditions / tests”.
Note:
If any of
the alarms fail to operate,
DO NOT USE
until the fault has been rectified.
7. Set EPAP to the prescribed pressure.
8. The patient may now
hold
the mask to the face.
9. Allow the patient to get used to the mask. Then fasten the interface to the patient.
10. Lock the settings to prevent unauthorised adjustment. See back page.
11. Remove the patient interface or disconnect the patient circuit from the outlet port,
before switching off.