Inlet
Power Cord
Jack
Power switch
Power lamp
Caution
4) Remove the contra head from the motor
handpiece and mount the motor cap onto it.
(Refer to "7. MOTOR CAP") Insert the motor
handpiece into the battery charger with the
operation panel facing frontward. If the motor
handpiece power is on, it will be turned off
automatically when inserted into the battery
charger. Charging starts with the charge lamp
turned on. The battery lamp of the motor
handpiece lights. (Fig. 6)
5) Charge at full capacity is completed, when the
charge lamp goes out. (The battery lamp
remains to lighted.) If the motor handpiece is
inserted into the battery charger with the power
on, the motor handpiece will be automatically
powered on when removed from the charger.
4. OPERATION PROCEDURES
1) Insert the power cord jack into the inlet at the
back of the battery charger. (Fig. 4)
2) Insert the power cord plug into the outlet of
the correct voltage.
3) Turn on the power switch. At this time, check
that the power lamp lights. (Fig. 5)
(1) Charging Batteries
Fig. 4
Fig. 5
Fig. 6
If the power lamp or charge lamp flashes,
turn on the power switch once again. If
they still flash, contact Brasseler USA.
Charge lamp
Battery lamp
Caution
· Connect to AC120V power supply only. Voltage variation within ±10% is permissible.
Voltage beyond this range may result in damage.
· Never use the battery charger for anything other than the motor handpiece of the Hygiene
Pro Cordless.
· The charging normally takes approx. 90 minutes, but it depends on battery conditions,
battery freshness, ambient temperature, etc. Older batteries are especially prone to
significantly shorter charging and operating times.
· Batteries may slightly heat up during charging, but this is not a failure. If the handpiece is
inserted or removed into/from the charger at short intervals, (approx. 5 minutes) charging
cannot be properly completed and by touching the battery part you may feel heat.
We recommend that you charge batteries for as long a cycle as possible.
· If the motor handpiece is placed on the charger, it will not automatically power on if left on
the charger for more than 3 hours.
· The power of the motor handpiece should not be turned on immediately after removing
from the charger, wait approx. 2 seconds or longer with the power off.