Inlet

Power Cord 

Jack

Power switch

Power lamp

Caution

4) Remove the contra head from the motor 

handpiece and mount the motor cap onto it. 
(Refer to "7. MOTOR CAP") Insert the motor 
handpiece into the battery charger with the 
operation panel facing frontward. If the motor 
handpiece power is on, it will be turned off 
automatically when inserted into the battery 
charger. Charging starts with the charge lamp 
turned on. The battery lamp of the motor 
handpiece lights. (Fig. 6)

5) Charge at full capacity is completed, when the 

charge lamp goes out. (The battery lamp 
remains to lighted.) If the motor handpiece is 
inserted into the battery charger with the power 
on, the motor handpiece will be automatically 
powered on when removed from the charger.

4. OPERATION PROCEDURES

1) Insert the power cord jack into the inlet at the 

back of the battery charger. (Fig. 4)

2) Insert the power cord plug into the outlet of 

the correct voltage.

3) Turn on the power switch. At this time, check 

that the power lamp lights. (Fig. 5)

(1) Charging Batteries

Fig. 4

Fig. 5

Fig. 6

If the power lamp or charge lamp flashes, 
turn on the power switch once again. If 
they still flash, contact Brasseler USA.

Charge lamp

Battery lamp

Caution

 · Connect to AC120V power supply only. Voltage variation within ±10% is permissible. 

Voltage beyond this range may result in damage.

 · Never use the battery charger for anything other than the motor handpiece of the Hygiene 

Pro Cordless.

 · The charging normally takes approx. 90 minutes, but it depends on battery conditions, 

battery freshness, ambient temperature, etc. Older batteries are especially prone to 
significantly shorter charging and operating times.

 · Batteries may slightly heat up during charging, but this is not a failure. If the handpiece is 

inserted or removed into/from the charger at short intervals, (approx. 5 minutes) charging 
cannot be properly completed and by touching the battery part you may feel heat. 
We recommend that you charge batteries for as long a cycle as possible.

 · If the motor handpiece is placed on the charger, it will not automatically power on if left on 

the charger for more than 3 hours.

 · The power of the motor handpiece should not be turned on immediately after removing 

from the charger, wait approx. 2 seconds or longer with the power off.

Summary of Contents for HygienePro Air

Page 1: ...afety be sure to follow them Cautions for handling and operation Classification DANGER WARNING CAUTION NOTICE Degree of Danger or Damage and Seriousness Explains an instruction where death or serious...

Page 2: ...efully the instruction manual by the battery manufacturer before use Do not use a bent damaged deformed or non ISO conforming bur as this may result in personal injury due to its sudden breaking or fl...

Page 3: ...es may recover after repeating complete discharge and full charge a few fimes If you do not use the product for a long period of time remove the batteries from the motor handpiece NOTICE Compact body...

Page 4: ...Fig 1 3 Packaging Contents Battery charger Motor handpiece Power cord F type spray nozzle Motor cap Contra head DPA 4 For Disposable Prophy Angles 4 1 Reduction...

Page 5: ...rns on the power and re pressing and holding the switch for 2 seconds or longer turns off the power In addition by pressing the switch when the power is on the rotation speed can be selected Power Rot...

Page 6: ...Fig 6 If the power lamp or charge lamp flashes turn on the power switch once again If they still flash contact Brasseler USA Charge lamp Battery lamp Caution Connect to AC120V power supply only Volta...

Page 7: ...l only turn off when the batteries are drained to a certain low voltage level Auto Power Off Function Fig 7 If the Battery lamp flashes at no load during rotation this means that the battery is at a l...

Page 8: ...hey click When removing the contra head pull it straight out Fig 8 3 Changing Contra Head Make certain that the contra head is securely assembled to the motor handpiece Motor handpiece Positioning pin...

Page 9: ...ave pouch to keep it clean until you use it Sterilize the contra head only For sterilization we recommend the autoclave sterilization method Sterilization is required first time you use and after each...

Page 10: ...g may be different nickel metal hydride batteries 600 700 mAh nominal 1 2V Or Ni Cad batteries 200 300 mAh nominal 1 2V However please take note that the operating time shortens because of their capac...

Page 11: ...move old batteries 6 Insert new batteries according to the polarity marking in the battery box A mix up between the plus and minus sides will not allow operation of the handpiece 7 Attach the battery...

Page 12: ...w the discharge completion voltage Battery drain The battery voltage dropped below the voltage near discharge completion Temporary voltage dropped due to battery drain or heavy loading Non operational...

Page 13: ...tteries Turn on the power Recharge the batteries Recharge the batteries Insert and remove the motor handpiece into from the battery charger several times Turn on the power of the battery charger Inser...

Page 14: ...ByYour Side in Dentistry One Brasseler Boulevard Savannah GA 31419 To order call 800 841 4522 or fax 888 610 1937 Visit our website www BrasselerUSA com 08 08 005 M...

Reviews: