background image

F type spray 

nozzle

Contra head

Spray

Fig. 9

Fig. 8

Caution

5. Cleaning

Lubricate the contra head only.
Lubricate it before each autoclaving or at least once a day, using spray lubricant. 
Screw the F type spray nozzle onto the spray nozzle by approx. 10 turns. Insert the 
F type spray nozzle into the rear part of the contra head and lubricate the head for 
2-3 seconds. (Fig. 9)

(1) Lubricating Contra Head

The contra head can be connected with the 
motor handpiece at 6 adjustable head 
positions. Align the positioning pins of the 
contra head with the positioning slots of the 
motor handpiece and insert the head until 
they click. When removing the contra head, 
pull it straight out. (Fig. 8)

(3) Changing Contra Head

Make certain that the contra head 
is securely assembled to the motor 
handpiece.

Motor handpiece

Positioning pin

Positioning slots

Contra head

Caution

When the motor handpiece becomes dirty, wipe it off with a cotton cloth moistened with 
rubbing alcohol.

(2) Cleaning Motor Handpiece

 

• 

Do not lubricate the motor handpiece.

 

• 

Before mounting the lubricated contra head to the motor handpiece, wipe off extra oil. 

Stand on its end or lean it in the proper direction for draining. Mount it after excess oil 
has been completely drained.

 

• 

Hold the contra head securely to prevent it from flying off by the pressure of the spray.

 

• 

Hold the spray upright.

 

• 

When cleaning the motor handpiece, do not use any solvent such as benzine or thinner.

Summary of Contents for HygienePro Air

Page 1: ...afety be sure to follow them Cautions for handling and operation Classification DANGER WARNING CAUTION NOTICE Degree of Danger or Damage and Seriousness Explains an instruction where death or serious...

Page 2: ...efully the instruction manual by the battery manufacturer before use Do not use a bent damaged deformed or non ISO conforming bur as this may result in personal injury due to its sudden breaking or fl...

Page 3: ...es may recover after repeating complete discharge and full charge a few fimes If you do not use the product for a long period of time remove the batteries from the motor handpiece NOTICE Compact body...

Page 4: ...Fig 1 3 Packaging Contents Battery charger Motor handpiece Power cord F type spray nozzle Motor cap Contra head DPA 4 For Disposable Prophy Angles 4 1 Reduction...

Page 5: ...rns on the power and re pressing and holding the switch for 2 seconds or longer turns off the power In addition by pressing the switch when the power is on the rotation speed can be selected Power Rot...

Page 6: ...Fig 6 If the power lamp or charge lamp flashes turn on the power switch once again If they still flash contact Brasseler USA Charge lamp Battery lamp Caution Connect to AC120V power supply only Volta...

Page 7: ...l only turn off when the batteries are drained to a certain low voltage level Auto Power Off Function Fig 7 If the Battery lamp flashes at no load during rotation this means that the battery is at a l...

Page 8: ...hey click When removing the contra head pull it straight out Fig 8 3 Changing Contra Head Make certain that the contra head is securely assembled to the motor handpiece Motor handpiece Positioning pin...

Page 9: ...ave pouch to keep it clean until you use it Sterilize the contra head only For sterilization we recommend the autoclave sterilization method Sterilization is required first time you use and after each...

Page 10: ...g may be different nickel metal hydride batteries 600 700 mAh nominal 1 2V Or Ni Cad batteries 200 300 mAh nominal 1 2V However please take note that the operating time shortens because of their capac...

Page 11: ...move old batteries 6 Insert new batteries according to the polarity marking in the battery box A mix up between the plus and minus sides will not allow operation of the handpiece 7 Attach the battery...

Page 12: ...w the discharge completion voltage Battery drain The battery voltage dropped below the voltage near discharge completion Temporary voltage dropped due to battery drain or heavy loading Non operational...

Page 13: ...tteries Turn on the power Recharge the batteries Recharge the batteries Insert and remove the motor handpiece into from the battery charger several times Turn on the power of the battery charger Inser...

Page 14: ...ByYour Side in Dentistry One Brasseler Boulevard Savannah GA 31419 To order call 800 841 4522 or fax 888 610 1937 Visit our website www BrasselerUSA com 08 08 005 M...

Reviews: