WARNING
• Do not handle the power cord with a wet hand. Doing so may result in an electric shock.
•
Do not spill water or a chemical solution onto or into the motor handpiece or battery
charger. Doing so may result in fire or electric shock due to a short-circuit or breakage
due to rust formation.
• Do not disassemble or alter the motor handpiece.
• Do not drop the motor handpiece or battery charger. Place the battery charger on a flat and
stable place.
• Should the battery fluid leak and get into your eyes, immediately wash eyes thoroughly with
clean water and see your doctor. Failure to do so may result in loss of sight.
• Should the battery fluid leak and adhere to skin or clothing, immediately wash the exposed
skin thoroughly with clean water and completely wash away the fluid. Failure to do so may
result in skin trouble.
• If you should notice a battery fluid leak within the motor handpiece, deformation of the
motor-handpiece casing or partial discoloring, immediately stop use and contact Brasseler USA.
• If the product will not be used for a long period of time, remove the batteries to avoid fluid leak.
• Exercise sufficient care in using the product by giving patient safety first priority
• The product is to be used for only dental treatment by qualified personnel.
• Use commercially available batteries recommended by Brasseler USA.
Read carefully the instruction manual by the battery manufacturer before use.
• Do not use a bent, damaged, deformed or non-ISO-conforming bur, as this may result in personal
injury due to its sudden breaking or flying off during rotation.
• Do not use or leave the product in a high-temperature environment such as under strong direct
sunlight, in a car under a blazing sun, by a fire, or near a stove.
• Check the product before use by rotating it, paying attention to looseness, vibration, noise and
temperature (heat generation). If any abnormal condition is found even slightly at that time,
immediately stop use and contact Brasseler USA.
• Always clean the shank of the bur before installing. Allowing dirt to enter the chuck could cause
loss of concentricity and deterioration of chucking force.
• Before changing the head or bur, turn off the power of the motor handpiece.
Changing with the power on may cause unintended rotation by accidental touching of the
ON/OFF switch.
• When inserting the motor handpiece into the battery charger, position the handpiece correctly.
Pushing it into the charger forcibly in the incorrect position may cause damage.
• Be careful not to drop conductive parts such as wires and safety pins into the charging terminal
area of the battery charger.
• Before inserting the motor handpiece into the battery charger, wipe it clean. If the charging
terminal area becomes dirty, rust formation may prevent charging, or result in short-circuit.
• Do not lubricate the motor handpiece.
• Do not heat sterilize the motor handpiece. Do not autoclave.
CAUTION