Operation
BOWA-IFU-13120-ERGOECO-S0-EN-20220420 Instructions for Use ERGOeco
15
6. Release the switch.
The tissue is coagulated.
The tissue has been grasped by the jaws and is coagulated.
7.4 Removal
1. Close the jaws by closing the handle.
2. Withdraw the instrument from the trocar sleeve.
7.5 After use
7.5.1 Spare parts
For an overview of the available spare parts, please refer to chapter 1.1.2 “Spare parts”.
To order spare parts, contact your BOWA dealership or visit our website:
www.bowa-medical.com.
WARNING
Incomplete coagulation!
Never use the AUTOSTART function.
Do not switch on the HF current before the jaws are in contact with
the tissue to be coagulated and ensure that they are within the
surgical field of view.
Inadvertent activation of the instrument can cause injuries to the patient.
WARNING
Incomplete coagulation due to soiled jaws!
Regularly clean the jaws with a damp cloth or a sterile plastic brush
if tissue adheres to them.
WARNING
Risk of patient injury due to fractured or damaged parts!
Check the instrument after each use. All parts must be present.
WARNING
Defective, worn or soiled jaws can cause instrument malfunction!
Clean soiled jaws after use with a damp cloth or a sterile plastic
brush. See also “Device-specific information”, page 17.
Inserts with damaged jaws must be replaced.