Operation
BOWA-IFU-13120-ERGOECO-S0-EN-20220420 Instructions for Use ERGOeco
13
7 Operation
7.1 Before use
The instrument is now assembled (see chapter 6 “Assembly”)
.
1. Switch on the HF device and connect the HF cable to the HF device.
2. Set the required settings on the HF device.
3. Always perform a thorough visual inspection and function check before each use of the
laparoscopic instrument (see see chapter 9.3 “Check for damage”).
7.2 Function test in the operating theatre
1. Check whether the jaws can be easily opened and closed using the handle.
2. Check the activation by means of the foot switch.
The activation signal is emitted during activation of the foot switch.
WARNING
Risk of patient injury!
Only use compatible HF generators.
Select the settings of the HF device according to the requirements of
the intervention.
Only use suitable devices and accessories wherever possible.
Only use intact and sterilised devices.
WARNING
Risk of patient injuries from burns or explosion!
Avoid contact with flammable gases and liquids
(such as skin cleansers, disinfectants and anaesthetic gases).
Avoid direct skin contact with HF cables.
Skin-to-skin contact (e.g. between the patient's arms and body)
should be avoided, e.g. by inserting dry gauze.