Ambient conditions
20
Instructions for Use ERGOeco
BOWA-IFU-13120-ERGOECO-S0-EN-20220420
HF cable
1. Check the connector for damage and corrosion.
9.4 Sterilisation packaging
Before sterilisation, the instrument or its individual parts must be placed in suitable
disposable sterilisation packaging (single or double packaging) and/or a suitable
sterilisation container.
The packaging must meet the following requirements:
•
BS EN ISO 11607-1
•
Suitable for steam sterilisation (resistant to temperatures up to 137°C, sufficient steam
permeability)
•
Regular maintenance (sterilisation container)
10 Ambient conditions
10.1 Storage and transport
The original packaging is intended for storing non-sterile instruments prior to their first
use. Protect the packaging from moisture and both direct and indirect sunlight.
CAUTION
Brittle or defective insulation can burn the patient!
Replace instrument components if there is any damaged insulation.
The shipping box is not intended for storing the sterilised device.
The shipping box is not intended for storing the sterilised device.