Safety instructions
10
Instructions for Use ERGOeco
BOWA-IFU-13120-ERGOECO-S0-EN-20220420
3.2.1 Patients with pacemakers
Pacemaker malfunction or damage to the pacemaker can endanger the life of the patient
or result in irreversible injuries to the patient.
If you are treating a patient with a pacemaker, consult the cardiologist first before
performing HF surgery.
Set the pacemaker to a fixed demand frequency.
Make sure that the pacemaker does not directly contact the jaws.
Have a functioning defibrillator on hand.
Carry out a postoperative pacemaker check.
3.3 EMC information
Since medical electrical devices are subject to special EMC precautions, please comply
with the following information.
The BOWA accessory is only intended to be connected to BOWA-specified HF devices.
Using the accessory with medical devices from other manufacturers may result in
higher emission levels or reduced interference immunity.
NOTICE
Combining medical devices is safe only if
the lowest power rating (rated voltage/type of current, ...) is used for
all devices in the desired combination(s).
doing so is permitted according to the purpose and by the interface
specification of the devices used in the combination.
The Instructions for use and interface specifications of medical
devices used in combination must be strictly observed.